Gas chromatographs
are really rather simple. The apparatus consists of a pressurized
tank (cylinder) of carrier gas, a pressure regulator to control
the flow rate of the gas through the chromatograph, a sample inlet,
the column, a detector with associated electronics, some kind of
interface to the outside world such as a recorder, and a flow meter
to measure the flow rate of carrier gas. Chromatographs also provide
heating for the column, the sample inlet, and the detector. The
temperatures of these three components can usually be controlled
independently.
A gas chromatograph uses a flow-through narrow tube known as the
column, through which different chemical constituents of a sample
pass in a gas stream (carrier gas, mobile phase) at different rates
depending on their various chemical and physical properties and
their interaction with a specific column filling, called the stationary
phase. As the chemicals exit the end of the column, they are detected
and identified electronically. The function of the stationary phase
in the column is to separate different components, causing each
one to exit the column at a different time (retention time). Other
parameters that can be used to alter the order or time of retention
are the carrier gas flow rate, and the temperature.
The procedure involves
vaporizing a sample and sweeping it through a column with a moving
stream of gas termed the mobile phase or the carrier gas. The
gases are commonly supplied by compressed gas cylinders. The sample
is introduced into the injection port. The most common type of
analysis involves the injection of 1 to 3 micro liters of a liquid
sample into a heated inlet, either manually or by an automated
injection device. The injection port is interfaced to the column
where the actual separation takes place. In most cases capillary
columns are used to obtain the best possible separation. The downside
to capillary columns is that they have a limited "capacity".
In other words not much sample can be separated at a time. Therefore
a split injection port is often used to allow only a fraction
of the injected volume of sample onto the capillary column. The
capillary column's inner walls are coated with either a porous
solid or a viscous liquid material. This inner coating will interact
with different solute molecules to different extents. Those molecules
which interact more strongly with the stationary phase spend on
average a higher percentage of their time associated with the
stationary phase than those solutes which do not interact strongly.
Those compounds which do not often associate with the stationary
phase pass more quickly through the column than those compounds
which have strong interactions with the stationary phase, and
a separation of the components in the mixture is achieved. Since
the compounds have different mobilities, they exit the column
at different times; i.e., they have different retention times,
tR. The retention time is the time between injection and detection.
There are numerous detectors which can be used in gas chromatography.
It is a device that senses the presence of components different
from the carrier gas (mobile phase) and converts that information
to an electrical signal. For qualitative identification one must
rely on matching retention times of known compounds with the retention
times of components in the unknown mixture. It is important to
remember that any changes in operating conditions will affect
the retention time which will affect the accuracy of identification.
Thus GC is most often used when one is performing a target compound
analysis, where one has a good idea of the compounds present in
a mixture so reference standards can be used for determining retention
times. For a sample of largely unknown composition qualitative
identification can be determined by gas chromatography-mass spectrometry.
A mass spectrum of any or all peaks in the chromatogram is compared
with spectra contained in spectral libraries on the system's computer.
GC truly tells you is at which relative time
a component eluted from the column and that the detector was
sensitive to it. To make results meaningful, analysts need to
know which components at which concentrations are to be expected;
and even then a small amount of a substance can hide itself
behind a substance having both a higher concentration and the
same relative elution time. Last but not least it is often needed
to check the results of the sample against a GC analysis of
a reference sample containing only the suspected substance.
The modern gas chromatograph is a fairly complex instrument mostly
computer controlled. The samples are mechanically injected, the
analytical results are automatically calculated and the results
printed out, together with the pertinent operating conditions
in a standard format.
The major advantage of gas chromatography is the very high resolving
(i.e., separating) power provided by the capillary columns that
are typically used. A broad selection of stationary phases are
available that provide different retention characteristics giving
us many choices for our particular analytical application.
|